UTILITIES
Manual Part Number
141
Click this tool to turn on the Exablate Prostate system; when the device is on, the Device Status (in
the status bar) appears as Ready.
10.2.5. Connect/Disconnect to the Site Network
The MR system should be disconnected from the hospital network during a treatment. In case of an urgent
need to transfer data, use this option to connect/disconnect the MR to the site network.
1. Click this tool to connect to the hospital network.
2. As soon as the data transfer is completed, click this tool to disconnect the MR from the site
network and continue the treatment.
CAUTION
Do not treat a patient while the MR is connected to the hospital network.
[C-7]
10.2.6. Change Macro-Spot Planner Configuration
Different Prostate tissues may have a different response to the acoustic energy and as a result, respond
differently to various energy deposition regimens.
The Exablate Prostate system supports Two Planner Modes:
1. Fan: The Macro-spot planner adds the most energetic sub-spot first, while the energy per sub-spot
is reduced for further sub-spots, due to the simulated accumulated heat effect. For the same reason,
per each sub-spot, the most energetic focal point will be the most anterior one.
2. Uniform: The Macro-spot planner attempts to achieve uniform energy distribution within the ROT
both by geometrically spreading the energy in a uniform grid and by attempting to distribute energy
uniformly between all focal points.
Both Planner modes have an additional sub-mode (denoted by the suffix High), in which the planner will
attempt to reach a higher temperature threshold within the ROT.