2.9. Environmental Requirements
System Operating Conditions:
Atmospheric pressure: 700 to 1060 kPa
Altitude: -30 m (-100 ft.) to +3000 m (+9800 ft.)
Operator + Equipment room:
Temperature range: +15 to 32°C
Relative humidity: 30 to 70% (non-condensing)
Magnet room:
Temperature range: +15 to 24°C
Relative humidity: 30 to 60% (non-condensing)
System Storage & Transportation Conditions:
Atmospheric pressure: 700 to 1060 kPa
Temperature range: +5 to 40°C
Relative humidity: Up to 90%
Treatment Kit Storage & Transportation Conditions:
Storage conditions:
Temperature range: +14 to 30°C
Relative humidity: 40 to 80%
*Kit shall be stored in a shaded place.
Transportation conditions:
Temperature range: -30 to +60°C
Relative humidity: 15 to 90%