GETTING STARTED
Manual Part Number
50
Then, cover the rear part of the probe (starting from the neck of the probe and towards its cable) with
the probe cover drape.
8. Place the probe in the Preparation Station in the upright position.
9. Make sure a new water pouch is attached to the system.
10. On the Water System display, select (press on) Prepare.
11. Press the Fill button continuously, until the balloon is ~1/2 full.
12. Press on Start Circulation button to start water circulation. Verify there are no water leaks. In case leaks
are detected, restart the procedure using a new disposable shell.
NOTE:
If the water system buzzer sounds, press Mute on the water system control panel or Reset
to continue. Alternatively, you can press the blue button on the workstation console once
for mute and again to reset. See appendix B for details.
[N-7]
13. Press the Fill and Drain buttons alternately, waiting several seconds between cycles, until the balloon is
full and degassed. Make sure that there are no visible bubbles inside the balloon.
14. Bring the balloon to the desired insertion volume and press Insert on the water system’s display. This
will set the current water volume as the reference volume for insertion and extraction of the probe, and
will also pause the water circulation in the balloon.
15. Press on Start Circulation to resume water circulation. Verify that the water system is in Treat mode
before starting the treatment.