Additional Information Page 154
Fitting Prescription and related details can be adjusted on the
right hand side of the screen. Select your preferred fitting
prescription in the upper dropdown list.
Choose between Berger, DSL m[i/o], Half Gain, NAL-NL1,
NAL-NL2, NAL-R, NAL-RP, POGO1, POGO2, Third Gain, or
’Custom’ if you have edited your target with the Edit feature.
Target shown will be calculated based on the selected fitting
prescription and the audiogram and can be shown as REIG and/or
REAR targets. If no audiogram has been entered into the
audiogram screen, no targets will be displayed.
Note that fitting prescription settings (such as Age and Client type)
will differ depending on what fitting prescription is selected.
Measurement Details of the selected curve are displayed as a
table on the right hand side of the screen.
A Curve Comment for each curve can be typed into the comment
section on the right hand side.
Select a curve using the curve tag boxes under Curve display
options and write a comment in the comment section.
The comment will then appear in the comment section whenever
the curve is selected.
Curve Display Options are found in the lower right hand corner.
If you have measured more curves of the same type (e.g. REIG
curves), they will be listed by their input level. Tick the ones that are
to be displayed on the graph.
Right clicking on the input level in the curve display will give the
fitter various options.