Affinity
2.0
Additional Information Page 202
Horizontal axis (x-axis) at 6khz and must not be more negative than -5dB (BSA, 2007 & ISO, 2003).
12. You can now proceed to running the REIG measurements, there are 3 different intensity measurements
in this protocol but the main focus should be to fit the 65dB curve as this is the intensity for speech.
Please place the hearing devices into your patient’s ears along with the probe tubes for the following
measurements.
You will notice that there is a target in these measurements; in the next step it is shown how you can
amend the output of the patient’s hearing devices to meet these targets.
Note: the default view for the Binaural REM feature is to have the left and right aided measure displayed
side-by-side. However this can be swapped to an on-top comparison by pressing the button shown
below, there is an example of this view below.