Model 2110
5½ Digit Multimeter Reference Manual Section 5:
Remote commands
2110-901-01 Rev. C/August 2013 5-5
Common command reference
This topic provides a detailed reference for all common commands.
.
*CLS — clear status
This command clears the Status Byte Summary Register and all event registers.
*ESE <NRf> event enable command
This command enables bits in the Standard Event Enable Register. The selected bits are then reported to the
Status Byte.
Usage
*ESE 1
*SRE 32
Details
Use the *ESE command and the appropriate decimal value for each bit to set the enable register mask. Refer to
Standard Event Register (on page B-8) for information about values. For example, enable “operation complete”
by using the *ESE 1 command.
*ESE? event enable query
This query command queries the Standard Event Enable Register.
Usage
*ESE?
Details
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in the register.
Refer to Standard Event Register (on page B-8
) for information about values.
*ESR? event status register query
This query command queries the value of the Standard Event Status Register. This query command clears the
Standard Event Register.
Usage
*ESR?
Details
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in the register.
Refer to Standard Event Register (on page B-8
) for information about values.

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
BrandKeithley
Model2110
CategoryMultimeter
LanguageEnglish

Related product manuals