EasyManuals Logo

Keithley 2110 User Manual

Keithley 2110
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #139 background imageLoading...
Page #139 background image
Section
5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-32 2110-901-01 Rev. C/August 2013
MINimum
Set to minimum value.
MAXimum
Set to maximum value.
Query
command description
CALCulate:PERCent:TARGet?
Query the target value for percent math function.
CALCulate:PERCent:TARGet? MIN
Query the target minimum value for percent math function.
CALCulate:PERCent:TARGet? MAX
Query the target maximum value for percent math function.
CALCulate:STATe <b>
Turn the selected math function on or off.
Parameters
<b> description
0 or OFF
Turn selected math function on.
1 or ON
Turn selected math function off.
Query
command description
CALCulate:STATe?
Query the state of the selected math function. Returns 0 (off) or 1 (on).
DATA:FEED RDG_STORE,'CALCulate'|''
This command selects whether readings taken using the INITiate command are stored in the multimeter’s internal
memory (default) or not stored at all. In the default state (DATA:FEED RDG_STORE,'CALC'), up to 2000 readings
are stored in memory when INITiate is executed. The MEASure? And CONFigure commands automatically select
'CALC'. With memory disabled (DATA:FEED RDG_STORE,“”), readings taken using INITiate are not stored. This
may be useful with the averaging operation since it allows you to determine an average of the readings without
storing the individual values. An error will be generated if you attempt to transfer readings to the output buffer
using the FETCh? command.
parameter description
'CALC'
Stores up to 2000 readings in memory when INITiate is executed.
''
Readings taken using INITiate are not stored.
Query
command description
DATA:FEED?
Query the reading memory state. Returns “CALC” or “”.

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
Model2110
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
ContinuityYes
Diode TestYes
CategoryDigital Multimeter
Display5.5 digit
DC Current Range100 µA to 3 A
AC Current Range100 µA to 3 A
Resistance Range100 MΩ
Frequency Range3 Hz to 300 kHz
Frequency Accuracy0.01% + 1 count
Capacitance Range1 nF to 100 µF
Capacitance Accuracy1% + 5 counts
ConnectivityUSB, GPIB (optional)
Dimensions88 mm
Power Supply100 V to 240 V AC, 50/60 Hz
DC Voltage Accuracy±0.012% + 2 counts