odel 2110 5½ Digit Multimeter Reference Manual Section 5:
Remote commands
2110-901-01 Rev. C/August 2013 5-59
[SENSe:]TEMPerature:SPRTD:DX <n>|MIN|MAX
Use this command to set the DX coefficient for the SPRTD sensor.
Parameters
<n> description
-1.1 to +1.1
Select the DX coefficient for the SPRTD sensor.
MINimum
Select the minimum coefficient.
MAXimum
Select the maximum coefficient.
Query
command description
[SENSe:]TEMPerature:SPRTD:DX?
Query the sensor DX coefficient.
[SENSe:]TEMPerature:SPRTD:DX? MIN
Query the minimum sensor DX coefficient.
[SENSe:]TEMPerature:SPRTD:DX? MAX
Query the maximum sensor DX coefficient.
Details
Refer to the following topic in the Temperature measurement (on page 3-17) section for more information about
the DX coefficient: RTD coefficients and temperature calculation (on page 3-19).
[SENSe:]TEMPerature:TRANsducer <name>
Use this command to select the temperature transducer type.
Parameters
<name> description
RTD
Select 2-wire RTD.
FRTD
Select 4-wire RTD (also works with 3-wire RTD).
Query
command description
[SENSe:]TEMPerature:TRANsducer:TY
PE?
Query the temperature transducer type.
Details
Refer to the following topic in the Temperature measurement (on page 3-17) section for more information about
RTD measurement: RTD coefficients and temperature calculation (on page 3-19).

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
BrandKeithley
Model2110
CategoryMultimeter
LanguageEnglish

Related product manuals