5½ Digit Multimeter Reference Manual Section 5:
2110-901-01 Rev. C/August 2013 5-27
CALCulate subsystem
The commands in this subsystem configure and control the Calculate subsystems and are
summarized in CALCulate subsystem summary (on page 5-20
).
There are seven math operations. Only one of them can be enabled at a time. They either store data
for later use or perform mathematical operations on the readings. See Math operations (on page
3-
48) for more details.
CALCulate:AVERage:AVERage?
This command reads the average value of all readings taken since the averaging operation was enabled. The
instrument clears the value when averaging is turned on, when the power is turned off, or a remote interface reset
occurs.
CALCulate:AVERage:COUNt?
This command reads the average value of all readings taken since averaging was enabled. The instrument clears
the value when averaging is turned on, when the power is turned off, or a remote interface reset occurs.
CALCulate:AVERage:MINimum?
This command reads the minimum value found during the averaging operation. The instrument clears the value
when averaging is turned on, when the power is turned off, or a remote interface reset occurs.
CALCulate:AVERage:MAXimum?
This command reads the maximum value found during the averaging operation. The instrument clears the value
when averaging is turned on, when the power is turned off, or a remote interface reset occurs.