5½ Digit Multimeter Reference Manual Section 5:
Remote commands
2110-901-01 Rev. C/August 2013 5-63
[SENSe:]VOLTage[:DC]:NPLCycles <n>|MIN|MAX
Use this command to select the integration time for voltage measurements. Integration time for DCV
measurements is expressed as the number of power line cycles (NPLC).
Parameters
<n> description
0.001
Select 0.001 NPLC.
0.006
Select 0.006 NPLC.
0.02
Select 0.02 NPLC.
0.06
Select 0.06 NPLC.
0.2
Select 0.2 NPLC.
0.6
Select 0.6 NPLC.
1
Select 1 NPLC.
2
Select 2 NPLC.
10
Select 10 NPLC.
100
Select 100 NPLC.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]VOLTage[:DC]:NPLCycles?
Query the NPLCycles.
[SENSe:]VOLTage[:DC]:NPLCycles? MIN
Query the minimum value.
[SENSe:]VOLTage[:DC]:NPLCycles? MAX
Query the maximum value.
Details
Refer to Integration time and resolution for DC and resistance measurements (on page 3-41) more information
about NPLC.

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
BrandKeithley
Model2110
CategoryMultimeter
LanguageEnglish

Related product manuals