5: Remote commands Model 2110 5½ Digit Multimeter
5-56 2110-901-01 Rev. C/August 2013
[SENSe:]TEMPerature:SPRTD:RZERo <n>|MIN|MAX
Use this command to set the sensor R coefficient at 0 degrees Celsius.
Parameters
<n> description
Select sensor R coefficient at 0 degrees Celsius.
Select the minimum value.
Select the maximum value.
Query
command description
[SENSe:]TEMPerature:SPRTD:RZERo?
Query the sensor R value at 0 degrees Celsius.
[SENSe:]TEMPerature:SPRTD:RZERo? MIN
Query the minimum sensor R value at 0 degrees
[SENSe:]TEMPerature:SPRTD:RZERo? MAX
Query the maximum sensor R value at 0 degrees
Details
Refer to the following topic in the Temperature measurement (on page 3-17) section for more information about
SPRTD type: RTD coefficients and temperature calculation (on page 3-19).
[SENSe:]TEMPerature:SPRTD:A4 <n>|MIN|MAX
Use this command to set the A4 coefficient for the SPRTD sensor.
Parameters
<n> description
Select the minimum coefficient.
Select the maximum coefficient.
Query
command description
[SENSe:]TEMPerature:SPRTD:A4?
Query the sensor A4 coefficient.
[SENSe:]TEMPerature:SPRTD:A4? MIN
Query the minimum sensor A4 coefficient.
[SENSe:]TEMPerature:SPRTD:A4? MAX
Query the maximum sensor A4 coefficient.
Details
Refer to the following topic in the Temperature measurement (on page 3-17) section for more information about
the A4 coefficient: RTD coefficients and temperature calculation (on page 3-19).