Section
5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-50 2110-901-01 Rev. C/August 2013
[SENSe:]TCOuple:RJUNction:RSELect <name>
Use this command to select a reference junction type, real or simulated.
Parameters
<name> description
REAL
Select a real reference junction type.
SIMulated
Select a simulated reference junction type.
Query
command description
[SENSe:]TCOuple:RJUNction:RSELect?
Query the reference junction type, real or simulated.
Details
Refer to Thermocouple measurements (on page 3-22) for more information about sensor type.
[SENSe:]TCOuple:RJUNction:SIMulated <n>|MIN|MAX
Use this command to select the default temperature of the simulated reference junction.
Parameters
<n> description
-100 to 100
Select the default temperature value in degrees Celsius.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]TCOuple:RJUNction:SIMulated?
Query the temperature of the simulated reference
junction.
Details
Refer to the following topic in the Temperature measurement (on page 3-17) section for more information about
sensor type. Thermocouple measurements (on page 3-22).

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
BrandKeithley
Model2110
CategoryMultimeter
LanguageEnglish

Related product manuals