EasyManua.ls Logo

Keithley 2110

Keithley 2110
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 2110
5½ Digit Multimeter Reference Manual Section 5: Remote comman
ds
2110-901-01 Rev. C/August 2013 5-37
[SENSe:]CURRent:AC:RANGe <range>|MIN|MAX
Use this command to select the function (ACI) and set a measurement range.
Parameters
<n> description
0 to 10
Select the measurement range in amps.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]CURRent:AC:RANGe?
Query the measurement range.
[SENSe:]CURRent:AC:RANGe? MIN
Query the minimum range.
[SENSe:]CURRent:AC:RANGe? MAX
Query the maximum range.
[SENSe:]CURRent:AC:RESolution <n>|MIN|MAX
Use this command to select the resolution for AC current measurements (ACI).
Parameters
<n> description
(1.00e-06 to 1.00e-04) * ACI range
Select the resolution of the measurement.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]CURRent:AC:RESolution?
Query the measurement resolution.
[SENSe:]CURRent:AC:RESolution? MIN
Query the minimum resolution.
[SENSe:]CURRent:AC:RESolution? MAX
Query the maximum resolution.
Details
Refer to Bandwidth and resolution for AC measurements (on page 3-44) for more information about resolution.

Table of Contents

Other manuals for Keithley 2110

Related product manuals