EasyManua.ls Logo

Keithley 2110

Keithley 2110
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 2110
5½ Digit Multimeter Reference Manual Section 5:
Remote commands
2110-901-01 Rev. C/August 2013 5-7
*PSC <b> power-on status clear
This command enables or disables the power-on status clear function.
Usage
*PSC 0
*PSC 1
Details
When *PSC 1 is in effect, the Status Byte and Standard Event Register enable masks are cleared when power is
turned on. *PSC 1 is the default.
When *PSC 0 is in effect, the Status Byte and Standard Event Register enable masks are not cleared when
power is turned on.
The command is stored in nonvolatile memory.
*PSC? power-on status clear query
This query command returns the power-on status clear setting.
Usage
*PSC?
Details
Query the power-on status clear setting. Returns “0” (*PSC 0) or “1” (*PSC 1).
*RST reset command
This command resets the instrument to the power-on configuration.
Usage
*RST
Details
Reset the instrument to the power-on configuration. This command does not clear the error queue. Refer to
Power-on settings (on page 1-6
) for information about power-on configuration.

Table of Contents

Other manuals for Keithley 2110

Related product manuals