5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-58 2110-901-01 Rev. C/August 2013
[SENSe:]TEMPerature:SPRTD:BX <n>|MIN|MAX
Use this command to set the BX coefficient for the SPRTD sensor.
Parameters
<n> description
-1.1 to +1.1
Select the BX coefficient for the SPRTD sensor.
MINimum
Select the minimum coefficient.
MAXimum
Select the maximum coefficient.
Query
command description
[SENSe:]TEMPerature:SPRTD:AX?
Query the sensor BX coefficient.
[SENSe:]TEMPerature:SPRTD:AX? MIN
Query the minimum sensor BX coefficient.
[SENSe:]TEMPerature:SPRTD:AX? MAX
Query the maximum sensor BX coefficient.
Details
Refer to the following topic in the Temperature measurement (on page 3-17) section for more information about
the BX coefficient: RTD coefficients and temperature calculation (on page 3-19).
[SENSe:]TEMPerature:SPRTD:CX <n>|MIN|MAX
Use this command to set the CX coefficient for the SPRTD sensor.
Parameters
<n> description
-1.1 to +1.1
Select the CX coefficient for the SPRTD sensor.
MINimum
Select the minimum coefficient.
MAXimum
Select the maximum coefficient.
Query
command description
[SENSe:]TEMPerature:SPRTD:CX?
Query the sensor CX coefficient.
[SENSe:]TEMPerature:SPRTD:CX? MIN
Query the minimum sensor CX coefficient.
[SENSe:]TEMPerature:SPRTD:CX? MAX
Query the maximum sensor CX coefficient.
Details
Refer to the following topic in the Temperature measurement (on page 3-17) section for more information about
the CX coefficient: RTD coefficients and temperature calculation (on page 3-19).