3: Basic DMM operation Model 2110 5½ Digit Multimeter
Reference Manual
3-28 2110-901-01 Rev. C/August 2013
Remote operation
The following command is used to select a primary or secondary function:
FUNction[1|2] "function"
When you use the number one (1) or no number, the primary function is selected. When you use the
number two (2), the secondary function is selected. Here are some examples:
FUNCtion "VOLTage[:DC]" Selects DCV as primary function
FUNCtion1 "VOLTage[:DC]" Selects DCV as primary function
FUNCtion2 "VOLTage[:DC]" Selects DCV as secondary function
Use the following commands to select a primary or secondary function:
FUNCtion[1|2] "VOLTage[:DC]" DCV function
FUNCtion[1|2] "VOLTage:AC" ACV function
FUNCtion[1|2] "CURRent[:DC]" DCI function
FUNCtion[1|2] "CURRent:AC" ACI function
FUNCtion[1|2] "FREQuency" FREQ function
FUNCtion[1|2] "FREQuency[:VOLT]" FREQ function (voltage input)
FUNCtion[1|2] "FREQuency:CURR" FREQ function (current input)
FUNCtion[1|2] "PERiod" PERIOD function
FUNCtion[1|2] "PERiod[:VOLT]" PERIOD function (voltage input)
FUNCtion[1|2] "PERiod:CURR" PERIOD function (current input)
FUNCtion[1|2] "RESistance" 2 function
FUNCtion[1|2] "FRESistance" 4 function
FUNCtion[1|2] "CAPacitance" Capacitance function
FUNCtion[1|2] "TEMPerature" TEMP (RTD) function
FUNCtion[1|2] "TCOuple" Thermocouple function
FUNCtion[1|2] "NONE" Disable second function
FUNCtion[1|2]? Query the selected primary or secondary function
Triggering and data buffer
Triggering is used to initiate one or more measurements by the instrument. Up to 2000 readings can
be stored in the data buffer (memory).
You can control instrument trigger operation from the front panel or through a remote interface.

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
Model2110
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
ContinuityYes
Diode TestYes
CategoryDigital Multimeter
Display5.5 digit
DC Current Range100 µA to 3 A
AC Current Range100 µA to 3 A
Resistance Range100 MΩ
Frequency Range3 Hz to 300 kHz
Frequency Accuracy0.01% + 1 count
Capacitance Range1 nF to 100 µF
Capacitance Accuracy1% + 5 counts
ConnectivityUSB, GPIB (optional)
Dimensions88 mm
Power Supply100 V to 240 V AC, 50/60 Hz
DC Voltage Accuracy±0.012% + 2 counts