Model 2110
5½ Digit Multimeter Reference Manual Section 5:
Remote commands
2110-901-01 Rev. C/August 2013 5-31
CALCulate:MXB:MMFactor <n>|MIN|MAX
This command defines the m factor for the mX+b calculation.
Parameters
<n> description
-1e6 to 1e6
Set a value for m.
MINimum
Set value to the minimum.
MAXimum
Set value to the maximum.
Query
command description
CALCulate:MXB:MMFactor?
Query the value of m.
CALCulate:MXB:MMFactor? MIN
Query the minimum value of m.
CALCulate:MXB:MMFactor? MAX
Query the maximum value of m.
CALCulate:NULL:OFFSet <n>|MIN|MAX
This command stores a null value in the instrument’s Null Register. You must turn on the math operation before
writing to the math register.
Parameters
<n> description
0 to ± 120% of the
highest range
Select the null (relative) value. You can set the null value to any number
between -120 % and 120 % of the highest range.
MINimum
Set value to the minimum.
MAXimum
Set value to the maximum.
Query
command description
CALCulate:NULL:OFFSet?
Query the null value.
CALCulate:NULL:OFFS? MIN
Query the null minimum value for null math function.
CALCulate:NULL:OFFS? MAX
Query the null maximum value for null math function.
CALCulate:PERCent:TARGet <n>|MIN|MAX
This command specifies the target value for percent math function.
Parameters
<n> description
Specify the measurement function dependent target value.

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
Model2110
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
ContinuityYes
Diode TestYes
CategoryDigital Multimeter
Display5.5 digit
DC Current Range100 µA to 3 A
AC Current Range100 µA to 3 A
Resistance Range100 MΩ
Frequency Range3 Hz to 300 kHz
Frequency Accuracy0.01% + 1 count
Capacitance Range1 nF to 100 µF
Capacitance Accuracy1% + 5 counts
ConnectivityUSB, GPIB (optional)
Dimensions88 mm
Power Supply100 V to 240 V AC, 50/60 Hz
DC Voltage Accuracy±0.012% + 2 counts

Related product manuals