EasyManuals Logo

Keithley 2110 User Manual

Keithley 2110
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #175 background imageLoading...
Page #175 background image
Section
5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-68 2110-901-01 Rev. C/August 2013
TRIGger:COUNt <n>|MIN|MAX|INFinite
Use this command to specify the number of triggers the instrument will accept before returning to the “idle” state.
Select from 1 to 50,000 triggers. Trigger count is ignored while in local operation.
Parameters
<n> description
1 to 50,000
Select value for the number of triggers.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
INFinite
Select INFinite to instruct the instrument to take triggers continuously.
Query
command description
TRIGger:COUNt?
Query the number of triggers. Returns “9.90000000E+37” when you
select the infinite trigger count.
TRIGger:COUNt? MIN
Query the minimum number of triggers.
TRIGger:COUNt? MAX
Query the maximum number of triggers.
TRIGger:DELay <n>|MIN|MAX
Use this command to select a trigger delay time in seconds. The delay is the time between the trigger signal and
each sample that follows. Specify a delay time from 0 to 3600 seconds.
Parameters
<n> description
0 to 3600
Specify a delay time from 0 to 3600 seconds.
Query
command description
TRIGger:DELay?
Query the trigger delay time.
TRIGger:DELay? MAX
Query the minimum trigger delay time.
TRIGger:DELay? MIN
Query the maximum trigger delay time.
Details
Refer to Trigger delay (on page 3-34) for information about sources.

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
Model2110
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
ContinuityYes
Diode TestYes
CategoryDigital Multimeter
Display5.5 digit
DC Current Range100 µA to 3 A
AC Current Range100 µA to 3 A
Resistance Range100 MΩ
Frequency Range3 Hz to 300 kHz
Frequency Accuracy0.01% + 1 count
Capacitance Range1 nF to 100 µF
Capacitance Accuracy1% + 5 counts
ConnectivityUSB, GPIB (optional)
Dimensions88 mm
Power Supply100 V to 240 V AC, 50/60 Hz
DC Voltage Accuracy±0.012% + 2 counts

Related product manuals