5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-46 2110-901-01 Rev. C/August 2013
[SENSe:]PERiod:APERture <n>|MIN|MAX
Use this command to set the aperture time for period function. For frequency and period measurements, aperture
time is analogous to integration time.
Parameters
<n> description
0.01
-digit resolution.
0.1
-digit resolution.
1
-digit resolution.
Query
command description
[SENSe:]PERiod:APERture?
Query the aperture time.
[SENSe:]PERiod:APERture? MIN
Query the minimum aperture time.
[SENSe:]PERiod:APERture? MAX
Query the maximum aperture time.
Details
Refer to Aperture and resolution for frequency and period measurements (on page 3-45) for more information
about aperture time.
[SENSe:]PERiod:VOLTage:RANGe <n>|MIN|MAX
Use this command to select the voltage range for the period function.
Parameters
<n> description
0 to 750.
Set AC voltage range for the period function.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]PERiod:VOLTage:RANGe?
Query the measurement range.
[SENSe:]PERiod:VOLTage:RANGe? MIN
Query the minimum range.
[SENSe:]PERiod:VOLTage:RANGe? MAX
Query the maximum range.