Page #141 background imageLoading...
Page #141 background image
Section
5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-34 2110-901-01 Rev. C/August 2013
DISPlay:TEXT:CLEar
This command clears text messages from the secondary display.
Parameters
command description
DISPlay:Text:CLEar
Clears the message.
Example
DISPlay:TEXT:CLEar
Clears the message.
SENSe subsystem
The sense subsystem controls measurement configuration and is summarized in SENSe subsystem
summary (on page 5-21).
[SENSe:]AVERage:COUNt <n>|MIN|MAX
Use this command to set the number of readings the digital filter stores before calculating the average of the
stored readings. The digital filter averages the readings to yield a single reading.
Parameters
<n> description
2 to 100
Select the number of readings the digital filter stores before calculating the average.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]AVERage:COUNt?
Query the number of readings to average.
[SENSe:]AVERage:COUNt? MIN
Query the minimum number of readings to average.
[SENSe:]AVERage:COUNt? MAX
Query the maximum number of readings to average.
Details
Refer to Digital filter (on page 3-47) for more information about using this command.

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
Model2110
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
ContinuityYes
Diode TestYes
CategoryDigital Multimeter
Display5.5 digit
DC Current Range100 µA to 3 A
AC Current Range100 µA to 3 A
Resistance Range100 MΩ
Frequency Range3 Hz to 300 kHz
Frequency Accuracy0.01% + 1 count
Capacitance Range1 nF to 100 µF
Capacitance Accuracy1% + 5 counts
ConnectivityUSB, GPIB (optional)
Dimensions88 mm
Power Supply100 V to 240 V AC, 50/60 Hz
DC Voltage Accuracy±0.012% + 2 counts