5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-34 2110-901-01 Rev. C/August 2013
DISPlay:TEXT:CLEar
This command clears text messages from the secondary display.
Parameters
command description
DISPlay:Text:CLEar
Clears the message.
Example
DISPlay:TEXT:CLEar
Clears the message.
SENSe subsystem
The sense subsystem controls measurement configuration and is summarized in SENSe subsystem
summary (on page 5-21).
[SENSe:]AVERage:COUNt <n>|MIN|MAX
Use this command to set the number of readings the digital filter stores before calculating the average of the
stored readings. The digital filter averages the readings to yield a single reading.
Parameters
<n> description
2 to 100
Select the number of readings the digital filter stores before calculating the average.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]AVERage:COUNt?
Query the number of readings to average.
[SENSe:]AVERage:COUNt? MIN
Query the minimum number of readings to average.
[SENSe:]AVERage:COUNt? MAX
Query the maximum number of readings to average.
Details
Refer to Digital filter (on page 3-47) for more information about using this command.