EasyManua.ls Logo

Keithley 2110

Keithley 2110
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section
5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-66 2110-901-01 Rev. C/August 2013
SYSTem:IDNStr <string>
Use this command to change the identification string.
Parameters
<string> description
"MANUFACTURER,PRODUCT"
The identification string. No more than 39 characters.
Details
To query the identification string, use *IDN?. Make sure that the DMM is in compatible mode (see Selecting the
language (on page 2-23) for details).
SYSTem:LOCal
Place the instrument in the local mode. All keys on the front panel are fully functional.
Query
None.
SYSTem:REMote
Place the multimeter in the remote mode. All buttons on the front panel, except the LOCAL button, are disabled.
Query
None.
SYSTem:VERSion?
Query the present SCPI version.
Query
This query command returns the SCPI version. For example:
1991.0