Section
5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-22 2110-901-01 Rev. C/August 2013
Command Description
:NPLCycles <n>|MIN|MAX
Specify the integration time in number of power line cycles for
the selected function: 0.001, 0.006, 0.02, 0.06, 1, 2, 10, 100,
MIN, or MAX. Refer to
Integration time and resolution for DC and
resistance measurements (on page 3-41) for more information
about valid values for n.
:NPLCycles?[MIN|MAX]
Query NPLCycles.
:VOLTage:AC
Path to configure AC voltage.
:RANGe <n>|MIN|MAX
Specify range: 100 mV, 1 V, 10 V, 100 V, 750 V, MIN, or MAX.
? [MIN|MAX]
Query the range.
:AUTO <b>
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
:AUTO?
Query auto range.
:RESolution <n>|MIN|MAX
Specify resolution. Where <n> = 1.00e-06 to 1.00e-04 of range,
MIN, or MAX. Refer to Bandwidth and resolution for AC
measurements (on page 3-44) for more information about valid
values for n.
:RESolution?
Query the resolution.
:CURRent:AC
Path to configure AC current.
:RANGe <n>|MIN|MAX
Specify range:0 to 10, MIN, or MAX.
? [MIN|MAX]
Query the range for AC voltage.
:AUTO <b>
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
:AUTO?
Query auto range.
:RESolution <n>|MIN|MAX
Specify resolution. Where <n> = 1.00e-06 to 1.00e-04 of range,
MIN, or MAX. Refer to Bandwidth and resolution for AC
measurements (on page 3-44) for more information about valid
values for n.
:RESolution? [MIN|MAX]
Query the resolution.
:CURRent[:DC]
Path to configure DC current.
:RANGe <n>|MIN|MAX
Specify range: 10 mA, 100 mA, 1 A, 3 A, 10 A, MIN, or MAX.
? [MIN|MAX]
Query the range for DC voltage.
:AUTO <b>
Enable or disable auto range. Where: 0 = off and 1 = on.
:AUTO?
Query auto range.
:RESolution <n>|MIN|MAX
Specify resolution: <n> = 3.00e-07 to 3.00e-04 of range, MIN, or
MAX. Refer to Integration time and resolution for DC and
resistance measurements (on page 3-41) for more information
about valid values for n.
:RESolution ?
Query the resolution.
:NPLCycles <n> MIN|MAX
Specify the integration time in number of power line cycles for
the selected function: 0.001, 0.006, 0.02, 0.06, 1, 2, 10, 100,
MIN, or MAX. Refer to Integration time and resolution for DC and
resistance measurements (on page 3-41) for more information
about valid values for n.
:NPLCycles? [MIN|MAX]
Query NPLCycles.
:RESistance
Path to configure resistance (2-wire
).
:RANGe <n>|MIN|MAX
Specify range: 0 to 100e6, MIN, or MAX.
? [MIN|MAX]
Query resistance range.
:AUTO <b>
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
:AUTO?
Query auto range.
:RESolution <n>|MIN|MAX
Specify resolution: <n> = 3.00e-07 to 3.00e-04 of range, MIN, or
MAX. Refer to
Integration time and resolution for DC and
resistance measurements (on page 3-41) for more information
about valid values for n.
:RESolution? [MIN|MAX]
Query the resistance resolution.

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
Model2110
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
ContinuityYes
Diode TestYes
CategoryDigital Multimeter
Display5.5 digit
DC Current Range100 µA to 3 A
AC Current Range100 µA to 3 A
Resistance Range100 MΩ
Frequency Range3 Hz to 300 kHz
Frequency Accuracy0.01% + 1 count
Capacitance Range1 nF to 100 µF
Capacitance Accuracy1% + 5 counts
ConnectivityUSB, GPIB (optional)
Dimensions88 mm
Power Supply100 V to 240 V AC, 50/60 Hz
DC Voltage Accuracy±0.012% + 2 counts