Model 2110
5½ Digit Multimeter Reference Manual Section 3:
Basic DMM operation
2110-901-01 Rev. C/August 2013 3-21
Use the following commands to enter the parameter coefficients for the USER sensor:
TEMPerature:RTD:RZERo <n>
<n> = R-zero value
TEMPerature:RTD:ALPHa <n>
<n> = Alpha value
TEMPerature:RTD:BETA <n>
<n> = Beta value
TEMPerature:RTD:DELTa <n>
<n> = Delta value
Use the following commands to enter the parameter coefficients for the SPRTD sensor:
TEMPerature:SPRTd:RZERo <n>
<n> = R-zero value
TEMPerature:SPRTd:A4 <n>
<n> = A4 value
TEMPerature:SPRTd:B4 <n>
<n> = B4 value
TEMPerature:SPRTd:AX <n>
<n> = AX value
TEMPerature:SPRTd:BX <n>
<n> = BX value
TEMPerature:SPRTd:CX <n>
<n> = CX value
TEMPerature:SPRTd:DX <n>
<n> = DX value
Use the following commands to enter the parameter coefficients for the NTCT coefficient remote
commands.
TEMPerature:NTCT:A <n>
<n> = A value
TEMPerature:NTCT:B <n>
<n> = B value
TEMPerature:NTCT:C <n>
<n> = C value
Use the following command to measure temperature:
MEASure:TEMPerature?