List triggering 
You must configure list triggering before running a list. List triggering is similar to common triggering. 
However, list triggering introduces dwell time to control the source-delay-measure cycle. For more 
information about the common triggering, refer to Trigger model (on page 3-25). 
When running a list from the front panel, you only need to configure the arm source and a sample 
count for the trigger model. For other parameters, the instrument changes the settings according to 
the list configuration. 
Initiate continuous state 
Once the instrument finishes the list, it goes to idle mode. You can press the TRIGGER key to run 
the list again. 
When executing a list using SCPI commands, you need to set up the trigger model to match the list 
configuration. For details, refer to Configure and execute a 10-step linear list sweep (on page 5-3).