Models 2500 and 2502 User’s Manual Triggering 10-15
/ ↑↓STEST — Event detection occurs when the SOT line of the Digital I/O port
is pulsed either high or low. After selecting this arm event, you will be
prompted to select the state of the event detection bypass.
ARM OUT — Use to configure the arm layer output trigger:
/ LINE — Select the Trigger Link line for the output trigger: line #1, #2, #3, #4,
#5, or #6.
/ EVENTS — Enable (ON) or disable (OFF) the arm layer output triggers.
TRIG-LAYER-DONE ON enables an output trigger on exiting the trigger
layer.
COUNT — Specify the arm count, FINITE (programmable count) or INFINITE
(never ending count).
TRIG-LAYER — Use this menu item to configure the trigger layer of the trigger
model:
TRIGGER-IN — Use to select the detection event for the trigger layer:
/ IMMEDIATE — Event detection occurs immediately.
/ TRIGGER-LINK — After selecting this trigger-in source, you will be
prompted in sequence as follows:
> TRIG-IN TLINK LINE — Select the input line (#1, #2, #3, #4, #5, or #6)
for the Trigger Link.
> EVENT-DETECT-BYPASS — Set the bypass for the Source Event Detec-
tor. With ONCE, operation will loop around the Source Event Detector.
With NEVER selected, operation will wait for an input trigger.
> TRIGGER-IN-EVENTS — Enable (ON) or disable (OFF) trigger-in events
(SOURCE, DELAY, and MEAS). With a trigger-in event ON, operation
will wait at that event for an input trigger. With the trigger-in event OFF,
operation will not wait. It will simply continue and perform the appropriate
action.
TRIGGER OUT — Use to configure the trigger layer output trigger:
/ LINE — Select the Trigger Link line for the output trigger; line #1, #2, #3, #4,
#5, or #6.
/ EVENTS — Enable (ON) or disable (OFF) output triggers that occur after the
SOURCE, DELAY, and MEAS actions.
DELAY — Specify the time delay (in seconds) for the trigger delay.
COUNT — Specify the trigger count.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals