7-14 Status Structure Models 2510 and 2510-AT User’s Manual
Measurement event register
The used bits of the Measurement Event Register (shown in Figure 7-6) are described as
follows:
Bit B0, Over Temperature (OT) Set bit indicates an over temperature
condition.
Bit B1, Under Temperature (UT) Set bit indicates an under temperature
condition.
Bit B2, Over Voltage (OV) Set bit indicates an over voltage condition.
Bit B3, Current Overload (CO) Set bit indicates a current overload condition.
Bit B4, Over Resistance (OR) Set bit indicates an over resistance condition.
Bit B5, Under Resistance (UR) Set bit indicates an under resistance condition.
Bit B6, Reading Available (RAV) Set bit indicates that a reading was taken and
processed.
Bit B7, Reading Overflow (ROF) Set bit indicates that the reading exceeds the
selected measurement range of the Model 2510.
Bits B8 through B10 Not used.
Bit B11, Output Enable (OE) Set bit indicates that the Digital I/O port output
enable line is asserted.
Bit B12, Setpoint Tolerance (SPT) Set bit indicates that the setpoint is within
the programmed tolerance for the required number of readings.
Bit B13, Open Lead (OL) Set bit indicates that the temperature sensor leads are
open.
Bit B14, Shorted Lead (SL) Set bit indicates that the temperature sensor leads
are shorted.
Bit B15 Always zero.