10-32 SCPI Command Reference Models 2510 and 2510-AT User’s Manual
:B <n>
[:SENSe[1]]:TEMPerature:THERmistor:B <n> Program thermistor sensor B value
Parameters <n> = -10 to +10 thermistor B coefcient
MINimum -10
MAXimum +10
DEFault 2.33894e-4
Query :B? Query thermistor B coefcient
Description This command programs the thermistor temperature sensor B
coefcient.
:C <n>
[:SENSe[1]]:TEMPerature:THERmistor:C <n> Program thermistor sensor C value
Parameters <n> = -10 to +10 thermistor C coefcient
MINimum -10
MAXimum +10
DEFault 8.85983e-8
Query :C? Query thermistor C coefcient
Description This command programs the thermistor temperature sensor C
coefcient.
Solid-state sensor parameters
[:GAIN] <n>
[:SENSe[1]]:TEMPerature:VSS[:GAIN] <n> Program voltage sensor gain
[:SENSe[1]]:TEMPerature:ISS[:GAIN] <n> Program current sensor gain
Parameters :VSS <n> = 0 to 9.999e-2 VSS solid-state sensor gain (V/K)
MINimum 0
MAXimum 9.999e-2
DEFault 1.0e-2
:ISS <n> = 0 to 9.999e-4 ISS solid-state sensor gain (A/K)
MINimum 0
MAXimum 9.999e-4
DEFault 1.0e-6
Query [:GAIN]? Query solid-state sensor gain
Description These commands program the solid-state sensor gain values. :VSS sets
voltage type sensor gain, and :ISS sets type current sensor gain.