10-52 SCPI Command Reference Models 2510 and 2510-AT User’s Manual
Read version of SCPI standard
:VERSion?
:SYSTem:VERSion? Read SCPI version
Description This query command is used to read the version of the SCPI standard
being used by the Model 2510. Example code:
1996.0
The above response message indicates the version of the SCPI standard.
Reset timestamp
:RESet
:SYSTem:TIME:RESet Reset timestamp
Description This action command is used to reset the absolute timestamp to 0 sec-
onds. The timestamp also resets to 0 seconds every time the Model 2510
is turned on. Note that time information can be returned with the
:READ? and :MEAS? queries by specifying a time element with the
:FORMat:ELEMents command (see FORMat subsystem,
page 10-19).
2-wire/4-wire sense mode
:RSENse <b>
:SYSTem:RSENse <b> Control 4-wire/2-wire sensing
Parameters <b> = 1 or ON Enable 4-wire sensor sense mode
0 or OFF Disable 4-wire sensing (2-wire mode)
Query :RSENse? Query enabled/disabled 4-wire sense state
Description :RSENs enables or disables 4-wire sensing for RTD and thermistor tem-
perature transducers. 4-wire sensing should be used for RTD and ther-
mistor sensors with <1k resistance.