EasyManuals Logo
Home>Keithley>Measuring Instruments>2510

Keithley 2510 User Manual

Keithley 2510
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #188 background imageLoading...
Page #188 background image
Models 2510 and 2510-AT User’s Manual SCPI Command Reference 10-49
:POSetup
:SYSTem:POSetup <name> Program power-on defaults
Parameters <name> = RST Power-up to *RST defaults
PRESet Power-up to :SYSTem:PRESet defaults
SAV0 Power-up to setup stored at memory
location 0
SAV1 Power-up to setup stored at memory
location 1
SAV2 Power-up to setup stored at memory
location 2
SAV3 Power-up to setup stored at memory
location 3.
SAV4 Power-up to setup stored at memory
location 4.
Query :POSetup? Query power-on setup
Description This command is used to select the power-on defaults. With RST
selected, the instrument powers up to the *RST default conditions. With
PRES selected, the instrument powers up to the :SYStem:PRESet
default conditions. Default conditions are listed in the Table 10-1
through Table 10-10.
With the SAV0-4 parameters specied, the instrument powers-on to the
setup that is saved in the specied location using the *SAV command.
Select power line frequency setting
:LFRequency <freq>
:SYSTem:LFRequency <freq> Select line frequency
Parameters <freq> = 50 50Hz setting
60 60Hz setting
Query :LFRequency? Query line frequency selection.
Description Use this command to manually select the line frequency setting (50 or
60Hz). The line frequency setting should match the power line fre-
quency, or readings may be noisy.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2510 and is the answer not in the manual?

Keithley 2510 Specifications

General IconGeneral
BrandKeithley
Model2510
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals