Models 2510 and 2510-AT User’s Manual SCPI Command Reference 10-49
:POSetup
:SYSTem:POSetup <name> Program power-on defaults
Parameters <name> = RST Power-up to *RST defaults
PRESet Power-up to :SYSTem:PRESet defaults
SAV0 Power-up to setup stored at memory
location 0
SAV1 Power-up to setup stored at memory
location 1
SAV2 Power-up to setup stored at memory
location 2
SAV3 Power-up to setup stored at memory
location 3.
SAV4 Power-up to setup stored at memory
location 4.
Query :POSetup? Query power-on setup
Description This command is used to select the power-on defaults. With RST
selected, the instrument powers up to the *RST default conditions. With
PRES selected, the instrument powers up to the :SYStem:PRESet
default conditions. Default conditions are listed in the Table 10-1
through Table 10-10.
With the SAV0-4 parameters specied, the instrument powers-on to the
setup that is saved in the specied location using the *SAV command.
Select power line frequency setting
:LFRequency <freq>
:SYSTem:LFRequency <freq> Select line frequency
Parameters <freq> = 50 50Hz setting
60 60Hz setting
Query :LFRequency? Query line frequency selection.
Description Use this command to manually select the line frequency setting (50 or
60Hz). The line frequency setting should match the power line fre-
quency, or readings may be noisy.