EasyManuals Logo
Home>Keithley>Measuring Instruments>2510

Keithley 2510 User Manual

Keithley 2510
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
Models 2510 and 2510-AT User’s Manual Connections 2-5
Sense selection
4-wire connections minimize the detrimental effects of lead resistance by sourcing the test
current through one set of wires (OUTPUT and INPUT F terminals) and measuring the
voltage across the device through a second set of wires (S terminals).
OUTPUT sensing
Sensing though the OUTPUT terminals is automatic with no user programming required.
To use sensing with OUTPUT connections, simply make 4-wire connections to the ther-
moelectric cooler as shown in Figure 2-2. Note that it is recommended that you use 4-wire
OUTPUT connections whenever the thermoelectric cooler voltage or resistance must be
accurately known.
INPUT sensing
4-wire sensing may be used for both RTD and thermistor sensors, but sense selection is
not automatic. The sense mode must be manually set when choosing a sensor type with the
CONFIG T or CONFIG R menus. See “Configuring functions,” page 3-3, for details.
NOTE 4-wire connections and 4-wire sensing should be used with RTD and thermistor
sensors with resistances <1k
to avoid errors caused by lead resistance.
Sensing considerations
Figure 2-4 shows a comparison of 2- and 4-wire sensing methods. In these examples, the
INPUT terminals are connected to a thermistor type temperature sensor, but the consider-
ations for RTD sensors and OUTPUT connections to the TEC are similar.
2-wire sensing
The 2-wire connecting method is shown in Figure 2-4A. To determine the sensor resis-
tance, the test current (I) is forced through the F+ and F- test leads and the resistance of the
thermistor being measured (R
T
). The Model 2510 then measures the voltage across the
thermistor through the same F+ and F- connecting wires and computes the thermistor
resistance from the measured voltage (V
M
) and the current as follows:
The main problem with 2-wire connections is the wire resistance (R
W
). Since the test cur-
rent (I) causes a small but significant voltage drop across the wire resistances, the mea-
sured voltage (V
M
) will not be exactly the same as the voltage (V
T
) directly across the
thermistor, and considerable resistance and temperature measurement error can result.
R
T
V
M
I
--------=

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2510 and is the answer not in the manual?

Keithley 2510 Specifications

General IconGeneral
BrandKeithley
Model2510
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals