Models 2510 and 2510-AT User’s Manual SCPI Command Reference 10-29
:AUTO <b>
[:SENSe[1]]:TEMPerature:CURRent:AUTO <b> Enable/disable auto sensor current
Parameters <b> = 1 or ON Enable auto sensor current
0 or OFF Disable auto sensor current
Query :AUTO? Query auto sensor current state
Description This command enables or disables the auto sensor current mode. When
the auto current mode is enabled, the sensor current is set as follows:
Range Current
100Ω RTD or thermistor 2.5mA
1kΩ RTD or thermistor 833.3µA
10kΩ thermistor 100µA
100kΩ thermistor 33.3µA
:TRANsducer <name>
[:SENSe[1]]:TEMPerature:TRANsducer <name> Select temperature function sensor
Parameters <name> = THERmistor Select thermistor sensor
RTD Select RTD sensor
VSS Voltage type solid-state sensor
ISS Current type solid-state sensor
Query :TRANsducer? Query sensor type
Description This command select the temperature sensor for the temperature func-
tion. All four sensor types are available for use with the temperature
function.
RTD sensor parameters
:TYPE <name>
[:SENSe[1]]:TEMPerature:RTD:TYPE <name> Select RTD sensor type
Parameters <name> = PT385 Select PT385 RTD sensor
PT3916 Select PT3916 RTD sensor
PT100 Select PT100 RTD sensor
D100 Select D100 RTD sensor
F100 Select F100 RTD sensor
USER Select USER sensor
Query :TYPE? Query RTD sensor type
Description This command selects the RTD temperature sensor type for the temper-
ature and function. When USER is selected, α, β, and δ must also be
user programmed (see below).