10-30 SCPI Command Reference Models 2510 and 2510-AT User’s Manual
:RANGe <n>
[:SENSe[1]]:TEMPerature:RTD:RANGe <n> Set RTD resistance range
Parameters <n> = 0 to 1000 Set RTD resistance range
(100 or 1000)
MINimum 0
MAXimum 1000
DEFault 100
Query :RANGe? Query sensor resistance range
Description This command programs the temperature sensor resistance range. It
affects only RTD sensor types. Nominal ranges are: 100 and 1000.
:ALPHa <n>
[:SENSe[1]]:TEMPerature:RTD:ALPHa <n> Program RTD sensor alpha
Parameters <n> = 0 to 0.01 RTD α coefficient
MINimum 0
MAXimum 0.01
DEFault 0.00385
Query :ALPHa? Query RTD α coefficient
Description This command programs the RTD temperature sensor α coefficient for
the USER type RTD sensor.
NOTE See “Sensor coefficients,” page 4-14.
:BETA <n>
[:SENSe[1]]:TEMPerature:RTD:BETA <n> Program RTD sensor beta
Parameters <n> = 0 to 1 RTD β coefficient
MINimum 0
MAXimum 1
DEFault 0.111
Query :BETA? Query RTD β coefficient
Description This command programs the RTD temperature sensor β coefficient for
the USER type RTD sensor.