Models 2510 and 2510-AT User’s Manual SCPI Command Reference 10-31
:DELTa <n>
[:SENSe[1]]:TEMPerature:RTD:DELTa <n> Program RTD sensor delta
Parameters <n> = 0 to 5 RTD δ coefficient
MINimum 0
MAXimum 5
DEFault 1.507
Query :DELTa? Query RTD δ coefficient
Description This command programs the RTD temperature sensor δ coefficient for
the USER type RTD sensor.
Thermistor sensor parameters
:RANGe <n>
[:SENSe[1]]:TEMPerature:THERmistor:RANGe <n> Set thermistor resistance range
Parameters <n> = 0 to 3e5 Set thermistor resistance range
(100, 1e3, 1e4, 1e5)
MINimum 0
MAXimum 3e5
DEFault 1e4
Query :RANGe? Query sensor resistance range
Description This command programs the temperature sensor resistance range. It
affects only thermistor sensor types. Nominal ranges are: 100, 1e3, 1e4,
and 1e5.
:A <n>
[:SENSe[1]]:TEMPerature:THERmistor:A <n> Program thermistor sensor A value
Parameters <n> = -10 to +10 thermistor A coefficient
MINimum -10
MAXimum +10
DEFault 1.13030e-3
Query :A? Query thermistor A coefficient
Description This command programs the thermistor temperature sensor A coeffi-
cient.
NOTE See “Sensor coefficients,” page 4-14.