10-36 SCPI Command Reference Models 2510 and 2510-AT User’s Manual
Resistance function
[:AMPLitude] <n>
:SOURce[1]:RESistance[:LEVel][:IMMediate][:AMPLitude] <n> Program resistance value
Parameters <n> = 1 to 1.000e3Ω 100Ω thermistor
5 to 9.999e3Ω 1kΩ thermistor
50 to 8.000e4Ω 10kΩ thermistor
500 to 2.000e5Ω 100kΩ thermistor
1 to 250Ω 100Ω RTD
5 to 3.000e3Ω 1kΩ RTD
MINimum 1Ω (100Ω RTD or thermistor)
5Ω (1kΩ RTD or thermistor)
50Ω (10kΩ thermistor)
500Ω (100kΩ thermistor)
MAXimum 1.000e3Ω (100Ω thermistor)
9.999e3Ω (1kΩ thermistor)
8.000e4Ω (10kΩ thermistor)
2.000e5Ω (100kΩ thermistor)
250Ω (100Ω RTD)
3.000e3Ω (1kΩ RTD)
DEFault 100Ω (100Ω RTD or thermistor)
1e3Ω (1kΩ RTD or thermistor)
1e4Ω (10kΩ thermistor)
1e5Ω (100kΩ thermistor)
Query [:AMPLitude]? Query resistance amplitude
Description This command programs the amplitude (setpoint) for the resistance
function. The amplitude can be set over a range that depends on sensor
type. The value cannot exceed the protection limits.
:LCONstants
:SOURce[1]:RESistance:LCONstants[:GAIN] <n> Program proportional constant
:SOURce[1]:RESistance:LCONstants:DERvative <n> Program derivative constant
:SOURce[1]:RESistance:LCONstants:INTegral <n> Program integral constant
Parameters [:GAIN] <n> = 0 to 1e5 Set proportional constant value
MINimum 0
MAXimum 1e5
DEFault 20.0
:DERivative <n> = 0 to 1e5 Set derivative constant value
MINimum 0
MAXimum 1e5
DEFault 0.0
:INTegral <n> = 0 to 1e5 Set integral constant value
MINimum 0
MAXimum 1e5
DEFault 0.6