Models 2510 and 2510-AT User’s Manual Status Structure 7-15
Figure 7-6
Measurement event status
CO
(B3)
OV
(B2)
UT
(B1)
OT
(B0)
OR
(B4)
UR
(B5)
RAV
(B6)
ROF
(B7)
(B8)
(B9)
(B10)
SPT
(B12)
Measurement
Condition Register
(B9)
(B10)
Measurement Event
Register
(B10)
Measurement Event
Enable Register
&
&
&
&
&
&
OR
To Measurement
Summary Bit
(MSB) of Status
Byte Register.
(See Figure 7-1.)
RAV = Reading Available
ROF = Reading Overflow
OE = Output Enable Asserted
SPT = Setpoint Tolerance
OL = Open Lead
SL = Shorted Lead
OE
(B11)
&
&
&
&
OL
(B13)
SL
(B14)
(B15)
OE
(B11)
SPT
(B12)
OL
(B13)
SL
(B14)
(B15)
OE
(B11)
SPT
(B12)
OL
(B13)
SL
(B14)
(B15)
&
&
&
&
stat:meas:cond?
stat:meas?
stat:meas:enab <NRF>
stat:meas:enab?
&
OT
(B0)
UT
(B1)
OV
(B2)
CO
(B3)
OR
(B4)
UR
(B5)
RAV
(B6)
ROF
(B7)
(B8)
(B9)
(B8)
ROF
(B7)
RAV
(B6)
UR
(B5)
OR
(B4)
CO
(B3)
OV
(B2)
UT
(B1)
OT
(B0)
OT = Over Temperature
UT = Under Temperature
OV = Over Voltage
CO = Current Overload
OR = Over Resistance
UR = Under Resistance
& = Logical AND
OR = Logical OR