EasyManuals Logo
Home>Keithley>Measuring Instruments>2510

Keithley 2510 User Manual

Keithley 2510
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #167 background imageLoading...
Page #167 background image
10-28 SCPI Command Reference Models 2510 and 2510-AT User’s Manual
:RANGe <n>
[:SENSe[1]]:RESistance:RTD:RANGe <n> Set RTD resistance range
Parameters <n> = 0 to 1000 Set RTD resistance range (100 or 1000)
MINimum 0
MAXimum 1000
DEFault 100
Query :RANGe? Query sensor resistance range
This command programs the temperature sensor resistance range. It affects
only RTD sensor types. Nominal ranges are: 100 and 1000.
:RANGe <n>
[:SENSe[1]]:RESistance:THERmistor:RANGe <n> Set thermistor resistance range
Parameters <n> = 0 to 3e5 Set thermistor resistance range (100,
1e3, 1e4, 1e5)
MINimum 0
MAXimum 3e5
DEFault 1e4
Query :RANGe? Query sensor resistance range
Description This command programs the temperature sensor resistance range. It affects
only thermistor sensor types. Nominal ranges are: 100, 1e3, 1e4, and 1e5.
Temperature function
:CURRent <n>
[:SENSe[1]]:TEMPerature:CURRent <n> Set RTD or thermistor sensor current
Parameters <NRf> = 3e-6 to 2.5e-3 Set sensor current (3.3e-6, 1e-5,
3.33e-5, 1e-4, 8.333e-4, 2.5e-3)
MINimum 3e-6
MAXimum 2.5e-3
DEFault 2.5e-3 (100 range)
8.333e-4 (1k range)
1e-4 (10k range)
3.33e-5 (100k range)
Query :CURRent? Query sensor current
Description This command programs the temperature sensor (transducer) excitation
current. It affects only thermistor and RTD sensor types. Nominal sen-
sor currents are: 3.3e-6, 1e-5, 3.33e-5, 1e-4, 8.333e-4, and 2.5e-3.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2510 and is the answer not in the manual?

Keithley 2510 Specifications

General IconGeneral
BrandKeithley
Model2510
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals