10-28 SCPI Command Reference Models 2510 and 2510-AT User’s Manual
:RANGe <n>
[:SENSe[1]]:RESistance:RTD:RANGe <n> Set RTD resistance range
Parameters <n> = 0 to 1000 Set RTD resistance range (100 or 1000)
MINimum 0
MAXimum 1000
DEFault 100
Query :RANGe? Query sensor resistance range
This command programs the temperature sensor resistance range. It affects
only RTD sensor types. Nominal ranges are: 100 and 1000.
:RANGe <n>
[:SENSe[1]]:RESistance:THERmistor:RANGe <n> Set thermistor resistance range
Parameters <n> = 0 to 3e5 Set thermistor resistance range (100,
1e3, 1e4, 1e5)
MINimum 0
MAXimum 3e5
DEFault 1e4
Query :RANGe? Query sensor resistance range
Description This command programs the temperature sensor resistance range. It affects
only thermistor sensor types. Nominal ranges are: 100, 1e3, 1e4, and 1e5.
Temperature function
:CURRent <n>
[:SENSe[1]]:TEMPerature:CURRent <n> Set RTD or thermistor sensor current
Parameters <NRf> = 3e-6 to 2.5e-3 Set sensor current (3.3e-6, 1e-5,
3.33e-5, 1e-4, 8.333e-4, 2.5e-3)
MINimum 3e-6
MAXimum 2.5e-3
DEFault 2.5e-3 (100 range)
8.333e-4 (1k range)
1e-4 (10k range)
3.33e-5 (100k range)
Query :CURRent? Query sensor current
Description This command programs the temperature sensor (transducer) excitation
current. It affects only thermistor and RTD sensor types. Nominal sen-
sor currents are: 3.3e-6, 1e-5, 3.33e-5, 1e-4, 8.333e-4, and 2.5e-3.