Models 2510 and 2510-AT User’s Manual SCPI Command Reference 10-33
:OFFSet <n>
[:SENSe[1]]:TEMPerature:VSS:OFFSet <n> Program voltage sensor offset
[:SENSe[1]]:TEMPerature:ISS:OFFSet <n> Program current sensor offset
Parameters :VSS <n> = -999.999 to 999.999 VSS solid-state sensor offset (K)
MINimum -999.999
MAXimum 999.999
DEFault 0.0
:ISS <n> = -999.999 to 999.999 ISS solid-state sensor offset (K)
MINimum -999.999
MAXimum 999.999
DEFault 0.0
Query :OFFSet? Query solid-state sensor offset
Description These commands program the solid-state sensor offset values. :VSS sets
voltage type sensor offset, and :ISS sets current type sensor offset.