10-6 SCPI Command Reference Models 2510 and 2510-AT User’s Manual
[:SENSe[1]]
✓
:TEMPerature
:
RTD
:RANGe <n>
:RANGe?
Set RTD resistance range. <n> = 0 to 1000
(100,1000) | MIN | MAX | DEF
Query RTD resistance range.
100
:ALPHa <n> Program alpha coefficient for USER type.
<n> = 0 to 0.01 | MIN | MAX | DEF
0.00385
:ALPHa? Query alpha coefficient.
:BETA <n> Program beta coefficient for USER type.
<n> = 0 to 1 | MIN | MAX | DEF
0.111
:BETA? Query beta coefficient.
:DELTa <n> Program delta coefficient for USER type.
<n> = 0 to 5 | MIN | MAX | DEF
1.507
:DELTa? Query delta coefficient.
:THERmistor Program thermistor parameters.
:RANGe <n> Set thermistor resistance range. <n> = 0 to 3e5
(100, 1e3, 1e4, 1e5) | MIN | MAX | DEF
1e4
:RANGe? Query thermistor resistance range.
:A <n> Set A parameter value.
<n> = -10 to +10 | MIN | MAX | DEF
1.13030e-3
:A? Query A parameter value.
:B <n> Set B parameter value.
<n> = -10 to +10 | MIN | MAX | DEF
2.33894e-4
:B? Query B parameter value.
:C <n> Set C parameter value.
<n> = -10 to +10 | MIN | MAX | DEF
8.85983e-8
:C? Query C parameter value.
:VSS Set voltage type solid-state sensor parameters.
[:GAIN] <n> Program sensor gain (scale factor).
<n> = 0 to 9.999e-2 | MIN | MAX | DEF
1.0e-2
[:GAIN]? Query sensor gain.
:OFFSet <n> Program sensor offset.
<n> = -999.999 to 999.999 | MIN | MAX | DEF
0.0
:OFFSet? Query sensor offset.
:ISS Set current type solid-state sensor parameters.
[:GAIN] <n> Program sensor gain (scale factor).
<n> = 0 to 9.999e-4 | MIN | MAX | DEF
1.0e-6
[:GAIN]? Query sensor gain.
:OFFSet <n> Program sensor offset.
<n> = -999.999 to 999.999 | MIN | MAX | DEF
0.0
:OFFSet? Query sensor offset.
Table 10-4 (continued)
:SENSe[1] subsystem commands
Command Description Default parameter SCPI