Models 2510 and 2510-AT User’s Manual SCPI Command Reference 10-39
:LCONstants
:SOURce[1]:TEMPerature:LCONstants[:GAIN] <n> Program proportional constant
:SOURce[1]:TEMPerature:LCONstants:DERvative <n> Program derivative constant
:SOURce[1]:TEMPerature:LCONstants:INTegral <n> Program integral constant
Parameters [:GAIN] <n> = 0 to 1e5 Set proportional constant value
MINimum 0
MAXimum 1e5
DEFault 20.0
:DERivative <n> = 0 to 1e5 Set derivative constant value
MINimum 0
MAXimum 1e5
DEFault 0.0
:INTegral <n> = 0 to 1e5 Set integral constant value
MINimum 0
MAXimum 1e5
DEFault 0.6
Query [:GAIN]? Query proportional constant
:DERivative? Query derivative constant
:INTegral? Query integral constant
Description These commands program the temperature function PID loop constants,
which must be set to appropriate values for suitable control characteris-
tics. See Section 4, PID Control Concepts, for information on choosing
appropriate values.