Model 2657A High Power System SourceMeter® Instrument Reference Manual Appendix E: Status model
2657A-901-01 Rev. B/December 2012 E-11
Figure 152: Operation status trigger timer, trigger blender, and remote registers
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status Remote
Summary Register
+
status.operation.remote
Command Available (CAV)
To Remote Summary Bit (REM) in
Operation Status Summary Register
(status.operation)
.
.
.
Queue Not Empty
Command Queue
Prompts Enabled (PRMPT)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status Trigger
Timer Summary Register
+
status.operation.instrument.trigger_timer
To Trigger Timer Bit (TRGTMR) in
Operation Status Instrument
Summary Register
(status.operation.instrument)
Trigger Overrun (TRGOVR)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status Trigger
Timer Overrun Register
+
status.operation.instrument.trigger_timer.trigger_overrun
Timer 1 (TMR1)
Timer 2 (TMR2)
Timer 3 (TMR3)
Timer 4 (TMR4)
Timer 5 (TMR5)
Timer 6 (TMR6)
Timer 7 (TMR7)
Timer 8 (TMR8)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status Trigger
Blender Summary Register
+
status.operation.instrument.trigger_blender
To Trigger Blender Bit (TRGBLND) in
Operation Status Instrument
Summary Register
(status.operation.instrument)
Trigger Overrun (TRGOVR)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status Trigger
Blender Overrun Register
+
status.operation.instrument.trigger_blender.trigger_overrun
To Trigger Blender Bit (TRGBLND) in
Operation Status Trigger Overrun
Summary Register
(status.operation.trigger_overrun)
To Trigger Timer Bit (TRGTMR) in
Operation Status Trigger Overrun
Summary Register
(status.operation.trigger_overrun)
Blender 1 (BLND1)
Blender 2 (BLND2)
Blender 3 (BLND3)
Blender 4 (BLND4)

Table of Contents