Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-111
Positive transition effect on questionable event register
Questionable event Positive transition effect on questionable event register
Volts summary Sets B0 when an invalid volts measurement occurs
Amps summary Sets B1 when an invalid amps measurement occurs
Temperature summary Sets B4 when an invalid external temperature reading occurs
Humidity summary Sets B9 when an invalid relative humidity reading occurs
Ohms summary Sets B10 when an invalid ohms measurement occurs
Coulombs summary Sets B11 when an invalid coulombs measurement occurs
Sequence test summary Sets B12 when a sequence test is aborted
Figure 129: Questionable transition filter