5-26 Keysight E5260/E5270 User’s Guide, Edition 6
Front Panel Operations
Making Measurement
To set integration time
This procedure sets the number of averaging samples of the high-speed A/D
converter or the integration time of the high-resolution A/D converter. The meaning
of the setting value depends on the ADC type and ADC operation mode settings.
The value should be increased for reliable and accurate measurement, or decreased
for high speed measurement. Refer to Table 4-4 on page 4-13. The setup is effective
for all channels that use the same ADC type.
1. Press the
Shift
key and the
Integ
key. The setup message shown in Table 4-4 is
displayed. The message depends on the
Integ
and
AD
settings.
2. Press the arrow keys to set the N value for the message.
3. Press the
MeasCh
key if you are setting the value for measurement channels that
use another ADC type, and perform 2.
4. Press the
Enter
key to make the setup effective, or press the
Exit
key to cancel
changing the setup.