7-16 Keysight E5260/E5270 User’s Guide, Edition 6
If You Have a Problem
Error Codes
253 Program memory is full.
Maximum of 2000 programs or 40000 commands can be stored in the
program memory. Refer to “ST” command in Programming Guide.
254 Invalid input for a memory program.
The GPIB GET command (TRIGGER statement in HP BASIC) and an
external trigger input are not allowed in a memory program (between
the ST and END commands).
255 Maximum nesting level is eight.
Nesting (one program calling another) of a memory program must be
eight levels or less.
260 Data output buffer is full.
Maximum 34034 measurement data items can be stored in the data
output buffer.
270 Search source channel must be set.
Before triggering the search measurement or sending the LSSV, LSSI,
BSSV, or BSSI command to set the synchronous search source, send
the LSV, LSI, BSV, or BSI command to set the primary search source.
271 Search monitor channel must be set.
Before triggering the search measurement, send the LGV, LGI, BGV, or
BGI command to set the search monitor channel.
273 Search and sync output modes must be the same.
The primary search source channel and the synchronous source channel
must be different, and they must be set to the same output mode
(voltage or current).
274 Search sync source is overflow.
Set the search sources so that the same output range is set to both
primary and synchronous search sources.
275 Search target must be compliance value or less.
The search target value must be less than or equal to the compliance
value of the search monitor channel. Correct the search target value or
the compliance value.
276 Start and stop must be different.
Set different values for the search start and stop values.

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals