Keysight E5260/E5270 User’s Guide, Edition 6 7-17
If You Have a Problem
Error Codes
277 Step must be output resolution or more.
Set the search step value to the output resolution or more.
278 Search and sync channels must be different.
Set the search source and the synchronous source to different channels.
279 Search monitor mode must be compliance side.
Send the LGI/BGI command to set the voltage source search monitor
channel, or send the LGV/BGV command to set the current source
search monitor channel.
303 Excess voltage in MPSMU.
Voltage that exceeds maximum voltage at the present current range was
detected by a MPSMU. All output switches were set to OFF.
305 Excess current in HPSMU.
Current that exceeds maximum current at the present voltage range was
detected by a HPSMU. All output switches were set to OFF.
307 Unsupported module.
This module is not supported by this firmware version. Until you
update the firmware, use the Keysight E5260/E5270 with this module
removed.
310 Interlock open operation error. Initialized.
Initialization was automatically performed because the E5260/E5270
failed to set its output to 0 V when the interlock circuit was opened in
the high voltage condition. Any module may be defective. Perform
self-test.
311 ASU control cable was connected/disconnected.
The E5270B must be turned off when the Atto Sense and Switch Unit
(ASU) is connected/disconnected.
603 Sweep and pulse channels must be different.
Set the sweep source and the pulse source to different channels for the
staircase sweep with pulsed bias measurement (MM5).
610 Quasi-pulse source channel must be set.
Before triggering the quasi-pulsed spot measurement, send the BDV
command to set the quasi-pulse source.

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers: