7-18 Keysight E5260/E5270 User’s Guide, Edition 6
If You Have a Problem
Error Codes
620 TGP specified incorrect I/O port.
Specify trigger input for the Ext Trig In port, or trigger output for the
Ext Trig Out port by the TGP command. Refer to “TGP” command in
Programming Guide.
621 Specify trigger input port for PAX/WSX.
No trigger input port was specified for the PAX or WSX command.
Specify the trigger input port, or set the port as the trigger input port.
Refer to “TGP” command in Programming Guide to set trigger port.
622 Specify trigger output port for OSX.
No trigger output port was specified for the OSX command. Specify the
trigger output port, or set the port as the trigger output port. Refer to
“TGP” command in Programming Guide to set trigger port.
630 Incorrect polarity of search step value.
For the linear search measurement. The step value must be positive if
start<stop, or negative if start>stop.
631 Number of search steps must be 1001 or less.
For the linear search measurement. The number of search steps
between start and stop must be 1001 or less. This means the |step| value
must be |stop-start|/1001 or more.
632 Search measurement was aborted.
Search measurement was aborted by the automatic abort function.
640 Search limits must be range/20000 or more.
For the binary search measurement. The limit value for the search
target must be range/20000 or more. where range means the
measurement range actually used for the measurement.
650 Data format must be ASCII to get time data.
The time stamp function is not available for the binary data output
format. To use the time stamp function, set the data output format to
ASCII.
655 Cannot connect/disconnect series resistor.
The series resistor status cannot be changed in the high voltage state.
Set the output voltage or the voltage compliance to ±42 V or less to
connect or disconnect the series resistor.

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals