Keysight E5260/E5270 User’s Guide, Edition 6 2-21
Introduction
Measurement Modules
Table 2-13 HPSMU Output Voltage Ranges and Resolutions
Table 2-14 HPSMU Output Current Ranges and Resolutions
Range Output Value
Setting
Resolution
Maximum
Current
2 V 0 ≤ |V| ≤ 2 V
1 mV
a
a. 100 μV for the GPIB remote mode.
± 1000 mA
20 V 0 ≤ |V| ≤ 20 V 1 mV ± 1000 mA
40 V 0 ≤ |V| ≤ 40 V 2 mV ± 500 mA
100 V 0 ≤ |V| ≤ 100 V 5 mV ± 125 mA
200 V 0 ≤ |V| ≤ 200 V 10 mV ± 50 mA
Range Output Value
Setting
Resolution
Maximum
Voltage
100 nA 0 ≤ |I| ≤ 115 nA 5 pA ± 200 V
1 μA0 ≤ |I| ≤ 1.15 μA 50 pA ± 200 V
10 μA0 ≤ |I| ≤ 11.5 μA 500 pA ± 200 V
100 μA0 ≤ |I| ≤ 115 μA 5 nA ± 200 V
1 mA 0 ≤ |I| ≤ 1.15 mA 50 nA ± 200 V
10 mA 0 ≤ |I| ≤ 11.5 mA 500 nA ± 200 V
100 mA 0 ≤ |I| ≤ 50 mA 5 μA ± 200 V
50 mA < |I| ≤ 115 mA 5 μA ± 100 V
1 A 0 ≤ |I| ≤ 50 mA 50 μA ± 200 V
50 mA < |I| ≤ 125 mA 50 μA ± 100 V
125 mA < |I| ≤ 500 mA 50 μA± 40 V
500 mA < |I|
≤ 1 A 50 μA± 20 V