Front Panel Menu Reference SCPI Command
Not available
To generatea measurement trigger:
TRIG:ACQ
Alternatively, if the trigger sourceis BUS, you can also program a
*TRG or an IEEE-488 <get> command.
As previously discussed, a trigger can also be generated by an output transient, a digital pin, and
output voltage or current level, or a user-defined expression. If any of these systems are configured as
the trigger source, the instrument will wait indefinitely for the trigger signal. If the trigger does not
occur, you must manually return the trigger system to the idle state.The following commands return
the trigger system to the idle state:
Front Panel Menu Reference SCPI Command
Select Measure\Control.
Then selectthe Abort control.
ABOR:ACQ
Fetch the measurement
After a trigger is received and the measurement completes, the trigger system returns to the idle
state.
Once the measurement completes, FETCh queries can retrieve the most recent measurement data
without initiating a new measurement or altering the data in the measurement buffer.
Front Panel Menu Reference SCPI Command
Not available
To return RMS voltage & current:
FETC:VOLT:ACDC?
FETC:CURR:ACDC?
To return the high level of a pulse:
FETC:VOLT:HIGH?
FETC:CURR:HIGH?
To return the low level of a pulse:
FETC:VOLT:LOW?
FETC:CURR:LOW?
To return the maximum value:
FETC:VOLT:MAX?
FETC:CURR:MAX?
To return the minimum value:
FETC:VOLT:MIN?
FETC:CURR:MIN?
To return array data:
FETC:ARR:VOLT?
FETC:ARR:CURR?
FETC:ARR:POW?
4 Using the Advanced Power System
140 Keysight N6900/N7900 Series Operating and Service Guide