DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
100 of 180
16021 - Malfunctioning PMT1: Gain out of range
16022 - Malfunctioning PMT2: Gain out of range
Description: For a given high voltage, the low gain of PMT1 might prevent it from
reaching the correct signal level.
Cause 1: The exposure on the PLATE is not correct during the calibration.
Action: Do the Calibration for the CR 825/850 SYSTEM.
Cause 2: The alignment of the optics is not correct.
Action: Do:
adjustment for the FOLD MIRROR
adjustment of the GALVO ROTATION
Cause 3: The PMT malfunctioned.
Action: Install a new PMT. Use PMTs and COLLECTOR AY.
Cause 4: The PMT/DAS BOARD A5 malfunctioned.
Action: Do the checkout for the PMT/DAS BOARD A5.
Description: For a given high voltage, the low gain of PMT2 might prevent it from
reaching the correct signal level.
Cause 1: The exposure on the PLATE is not correct during the calibration.
Action: Do the Calibration for the CR 825/850 SYSTEM.
Cause 2: The alignment of the optics is not correct.
Action: Do:
adjustment for the FOLD MIRROR
adjustment of the GALVO ROTATION
Cause 3: The PMT malfunctioned.
Action: Install a new PMT. Use PMTs and COLLECTOR AY.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals