DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
96 of 180
16005 - Calibration Error: Routine did not receive image in the time
allowed
Cause 1: The exposure of the PLATE is not correct.
Action: 1. Use the DOSIMETER to check the exposure of the PLATE.
2. Check for scratches on the edges of the SCREEN.
Cause 2: The GALVO is not aligned correctly.
Action: Do:
adjustment of the GALVO ROTATION
checkout for the GALVO BOARD A4
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action: Do the checkout for the DIGITIZER BOARD A3.
Cause 3: The GALVO BOARD A4 malfunctioned.
Action: Do the checkout for the GALVO BOARD A4.