DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
67 of 180
11009 - DUPLEX CAM: Move failed
11010 - DUPLEX CAM: Software command failed on the MSC BD
11014 - DUPLEX CAM: Timeout Error
11015 - Move command failed because SLOW SCAN not at Home Position
11016 - MSC BOARD A1: Software Error for DUPLEX CAM
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Description: The DUPLEX CAM received a command to move that was not correct.
Cause 1: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The DUPLEX CAM did not move in the time allowed.
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause 1: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The MOTION SYSTEM CONTROL BOARD A1 did not provide the RTXC
TIMER. This TIMER determines the timeouts for the motion of the
DUPLEX CAM.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals